CareFusion Alaris SE User Manual page 15

Table of Contents

Advertisement

Getting Started
Check-In and Configuration (Continued)
Alternative Rate Accuracy Qualification Test (Continued)
1.
Fill solution container with clean tap water.
2.
Close AccuSlide flow regulator clamp on
then insert spike into solution container.
3.
Open AccuSlide flow regulator clamp and prime set.
Ensure that all air is expelled from set.
4.
Close AccuSlide flow regulator clamp.
5.
Place flask in middle of scale.
6.
Load set into instrument and close latch.
7.
Verify no fluid flow or drops falling in drip chamber.
8.
Plug instrument into a properly grounded
9.
Place output of set so it drips into flask. Do not let set rest
on flask.
10. Press
key to turn channel on.
POWER
11. Set primary infusion rate to
12. Set
to
mL.
VTBI
20
13. Ensure that instrument—both channels if dual channel—is
set to Pressure mode.
14. Press
RUN/HOLD
tubing is fully primed—approximately one minute.
15. Press
RUN/HOLD
1.
The factory default for the Monitoring Options mode is
Pressure.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
mL/h.
400
1
key to start primary infusion. Infuse until
key to stop infusion.
set and
80VCS
outlet.