Troubleshooting and Maintenance
To ensure that the system remains in good operating condition,
both regular and preventive maintenance inspections are
required. Refer to the applicable Alaris SE pump service
manual for detailed inspection instructions.
REGULAR INSPECTIONS
Procedure
Inspect for Damage:
Case
Connector
Keypad

Cleaning

Start-Up
Inspect and clean the Alaris SE pump as needed, typically
between patient uses.
use:
DO NOT
Solutions containing phosphoric acid (Foamy
aromatic solvents (such as naphtha or paint thinner),
chlorinated solvents
Toluene), ammonia, acetone, benzene, xylene, or alcohol,
other than as specified below.
Hard or pointed objects to clean any part of instrument.
ressurized sprays on instrument.
P
Acceptable cleaning solutions are as follows (use per
manufacturers' instructions):
bleach to nine parts water), Vesphene, Manu-Klenz, and warm
water.
1.
Unplug power cord from
1.
Excluding
bleach solution in water.
10%
116

Inspection Requirements

Frequency
Each usage
Each usage
Each usage
As required
Each usage
Cleaning
1
(such as Trichloroethane,
bleach solution (one part
10%
outlet before cleaning.
AC
WARNING
Failure to perform these
inspections can result in improper
instrument operation.
CAUTION
WARNING
Turn the instrument off and unplug
the power cord from
before cleaning. Do not spray fluids
1
),
Q&A
directly onto the instrument or into
any connector. Do not steam
,
MEK
autoclave, EtO sterilize, immerse the
instrument, or allow fluids to enter
the instrument case. Failure to
follow these instructions might result
in an electrical hazard.
CAUTION
The solutions/solvents identified as
NOT
surfaces of the instrument.
User Manual—Alaris SE Pump, Rev2.X
Inspection Requirements
Preventive maintenance
inspections should only be
performed by qualified service
personnel.
for anomalies