General Information
Definitions (Continued)
Optional Modes
Optional modes include Loading Dose, Dose Rate Calculator, Multi-Step, and Multi-Dose. Each of these
can be enabled or disabled. When disabled, they will not appear on the Options menu and cannot be
accessed by the clinician.
Dose Rate Calculator
Loading Dose
Multi-Dose
Multi-Dose Alert
Multi-Step
Rate, Maximum
Maximum infusion rate may be set from 0.1 mL/h to 999.9 mL/h in 0.1 mL/h increments.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Configurable Settings (Continued)
Allows clinician to enter the flow rate or the drug dose rate for a continuous infusion.
The system then calculates the alternate parameter based on the drug
concentration and if used, the weight.
Allows the clinician to set up an initial infusion rate for a specific volume, to be
followed automatically by a maintenance rate (primary settings) from the same
container. This is useful for delivering fluid challenges.
Allows 1 to 24 doses to be programmed at equally spaced intervals on the same
instrument over a 24-hour period. This mode allows delivery of multiple, equal
doses from the same IV container at regularly scheduled intervals. Within this
mode, a delayed start option allows the instrument to be programmed to delay the
infusion start for up to 8 hours.
When enabled, this feature alerts the clinician of the completion of each dose
delivered during a Multi-Dose program.
Allows a sequential drug delivery program (up to nine steps) to be set, delivering
volumes of fluid at different rates at each step. This allows the instrument
parameters to be set up once and deliver a step profile, eliminating the need to
change the rate and VTBI after each step of the infusion. The infusion may be
programmed in rate/volume or volume/time.
Configurable Settings
91