Check-In and Configuration (Continued)
Alternative Rate Accuracy Qualification Test (Continued)
16. Adjust height of instrument and/or fluid container to attain a
head height of
30 ±1
middle of pumping mechanism and fluid level in one of
following sources:
bag or vented bottle (vent closed on administration set)
drip chamber (unvented bottle with vent open on
administration set)
17. Zero reading on scale.
18. Verify that primary infusion rate is
19. Reset
to
VTBI
40
20. Press
RUN/HOLD
21. Instrument runs approximately
to complete delivery and then goes into
instrument within one second of its entering
22. Make a note of scale reading in grams.
23. Note expected volume, as identified on
24. Do not remove
80VCS
following conclusions is determined:
Instrument passes rate verification and calibration is
not needed.
Rate calibration number was changed and instrument
now passes verification.
Mechanism replacement is required.
10
inches/
centimeters between
76.2 ±2.5
400
mL and clear volume infused.
key to start primary infusion.
seconds (six minutes)
360
set from instrument until one of
mL/h.
mode. Stop
KVO
mode.
KVO
set insert.
80VCS
Check-In and Configuration
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230