Configurable Settings; Definitions - CareFusion Alaris SE User Manual

Table of Contents

Advertisement

General Information
The configurable options are set in the configuration mode by qualified service personnel. If the
configuration settings need to be changed from the factory default settings, refer to the Technical
Service Manual or contact CareFusion Technical Support for technical, troubleshooting, and preventive
maintenance information.
Air-in-Line
Air-in-Line Threshold (microliters)
Air-in-Line Accumulator
Air-in-Line Reset
Audio
Transition Tone (secondary to primary)
Volume
Configuration Name
Allows a 4-digit instrument ID label to appear in the lower LCD display.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230

Configurable Settings

Definitions

Sets the upper limit for a single bolus of air to pass without an alarm.
In other words, it is the amount of air allowed to pass through the air-
in-line detector before an air-in-line alarm sounds. One of four
different air-in-line detection settings can be selected: 50, 100, 200,
500 mcL.
Detects the presence of multiple air bubbles that are too small to be
detected by a single bolus AIL detection limit. The accumulator
feature, when enabled, looks for 10-15% of the downstream fluid
path to be air before giving an
The volume of air that trips the accumulated air detection alarm
varies based on the current setting for a single air bolus.
This allows the clinician to respond to an air-in-line alarm, assess its
clinical significance, and choose whether or not to continue infusion
without removing air. The reset feature allows only the current
bubble to proceed without tripping an alarm.
Provides an audible tone when the secondary
zero, to indicate the infusion has transitioned to a primary rate.
If all audio volume levels are enabled, the audio volume can be
adjusted for alarms, alerts, and
Medium, or Low. The audio volume indicator in the lower
display indicates the selected audio volume. The instrument
can be configured to enable only Medium and High, or only
High audio volume levels if desired.
Configurable Settings
ACCUMULATED AIR IN LINE
VTBI
tone to either High,
KVO
alarm.
reaches
LCD
87