Troubleshooting and Maintenance
2.
Verify that
RS-232
3.
Use a soft cloth dampened with warm water and a mild,
nonabrasive cleaning solution.
A soft-bristled brush may be used to clean narrow
areas.
Use light pressure when cleaning pressure transducer
and air-in-line detector areas of pumping channels.
4.
Routinely clean flow sensor by running warm water over it
while actuating slider, and then thoroughly dry it.
If the instrument shows evidence of damage in transit, notify
the carrier's agent immediately. Do not return damaged
equipment to the factory before the carrier's agent has
authorized repairs.
If the instrument fails to respond as described in this document
and the cause cannot be determined, do not use the
instrument. Contact qualified CareFusion service personnel.
If it is necessary to return the instrument for service, obtain a
return authorization number prior to shipment. Carefully
package the instrument (preferably in the original packaging),
reference the return authorization information, and return it to
the appropriate service or distribution center. CareFusion does
not assume any responsibility for loss of, or damage to,
returned instruments while in transit.
Technical support, service information, applications, and
manuals may be obtained by contacting a CareFusion
representative. When submitting any request for service,
include:
model number
a description of difficulty experienced
instrument settings
administration set/lot number
solution(s) used
message displayed at time of difficulty
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Cleaning (Continued)
connector is covered.

Service Information

Technical Support

Service Information
WARNING
The instrument case should only
be opened by qualified personnel
using proper grounding
techniques. Prior to performing
maintenance, disconnect the
Alaris SE pump from
During servicing, an instrument's
configuration settings might be
reset to the factory defaults.
Qualified hospital/facility
personnel are responsible for
checking in the instrument and
ensuring that the current
hospital-approved instrument
system configuration settings are
correct.
power.