General Information
Trumpet and Start-Up Curves (Continued)
Resistance Mode (Continued)
Resistance Mode Start-up at 25 mL/h (initial)
50
45
40
35
30
25
20
15
10
5
0
-5
0
20
Resistance Mode Trumpet Curve at 25 mL/h (initial)
15
15.6
10
6.1
5
2.3
0
-2.0
-5
-10
-8.4
-15
0
5
10
Resistance Mode Trumpet Curve at 25 mL/h (48 hr)
15
13.4
10
4.1
5
0.5
0
-3.6
-5
-9.9
-10
-15
0
5
10
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
40
60
80
100
Time (min)
5.0
3.2
1.3
-0.2
15
20
25
Observation Interval (min)
3.1
1.4
-0.5
-2.0
15
20
25
Observation Interval (min)
Resistance Mode Start-up at 999.9 mL/h (initial)
2000
1800
1600
1400
1200
1000
800
600
400
200
0
-200
0
120
Resistance Mode Trumpet Curve at 999.9 mL/h (initial)
15
10
3.0
5
0.1
-0.3
0
1.5
-0.9
-1.9
-5
-10
-15
0
30
Resistance Mode Trumpet Curve at 999.9 mL/h (24 hr)
15
10
5
1.2
0
-0.2
-2.7
-2.9
-3.3
-5
-3.9
-10
-15
30
0
Trumpet and Start-Up Curves
20
40
60
80
Time (min)
-0.6
-0.7
-1.1
-1.2
-1.5
5
10
15
20
Observation Interval (min)
-3.1
-3.1
-3.5
-3.4
-3.7
5
10
15
20
Observation Interval (min)
100
120
-0.7
-1.0
25
30
-3.2
-3.3
25
30
105