Dynamic Monitoring System
Downstream Occlusion Detection
When using the Adjustable Pressure monitoring mode, a
pressure alarm limit may be varied from
maximum configured pressure limit, in
When measured pressure exceeds this level, an
condition occurs.
DOWNSTREAM
When a Downstream Occlusion is detected, one of following
responses occurs:
•
If Auto Restart Plus feature is on, a Checking Line
message appears, along with an audible tone (see Auto
Restart Plus Feature for further details.)
•
If Auto Restart Plus feature is off, an
alarm occurs.
DOWNSTREAM
Auto Restart Plus Feature
The Auto Restart Plus feature provides the ability to
automatically continue an infusion if downstream pressure
measurements indicate that an occlusion condition has cleared
within a
-second Checking Line period (excluding High
40
Resistance Monitoring mode).
The Checking Line message and tone occur whenever a
pressure measurement exceeds the selected alarm threshold.
If the pressure falls to less than one-third of the alarm limit
within
seconds, normal flow resumes. The Adjustable
40
Pressure mode allows control of the pressure alarm limit. If the
condition is not cleared, the
occurs and infusion stops until manually restarted.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Monitoring Options—Pressure Mode
25
25
OCCLUSION
OCCLUSION DOWNSTREAM
mmHg to the
mmHg increments.
OCCLUSION
alarm
Monitoring Options—Pressure Mode
OPTIONS
A
B
A
OCCLUSION
DOWNSTREAM
600 mmHg
run
57