Prompts

Display
DOSE WOULD EXCEED
DOSE UNDER
DOSE NOT PERMITTED
TARGET WOULD EXCEED
BOLUS DOSE OVER
BOLUS DOSE UNDER
BOLUS DOSE NOT PERMITTED
WEIGHT OUTSIDE LIMIT
RATE NOT PERMITTED
Icon
Description and Troubleshooting Guide
The infusion rate has been set to a value which exceeds a Soft Alert. Check infusion
setting, to continue with infusion at set rate press the
OVERRIDE LIMIT by pressing the YES softkey. If OVERRIDE LIMIT is not required press
the NO softkey and adjust rate below Soft Alert.
The infusion rate has been set to a value which is under a Soft Alert. Check infusion
setting, to continue with infusion at set rate press the
OVERRIDE LIMIT by pressing the YES softkey. If OVERRIDE LIMIT is not required press
the NO softkey and adjust rate above Soft Alert.
The infusion rate has been set above a Hard Limit. Check infusion setting and adjust rate
to appropriate required rate.
The target has been set to a value which exceeds a Soft Alert. Check infusion setting,
to continue with infusion at set target press the
OVERRIDE LIMIT by pressing the YES softkey. If OVERRIDE LIMIT is not required press
the NO softkey and adjust rate below Soft Alert.
The bolus dose has been set to a value which exceeds a Soft Alert. Check the
bolus setting, to continue with the bolus press the
OVERRIDE LIMIT by pressing the YES softkey. If OVERRIDE LIMIT is not required press
the NO softkey and adjust dose below Soft Alert.
The bolus dose has been set to a value which is under a Soft Alert. Check the
bolus setting, to continue with the bolus press the
OVERRIDE LIMIT by pressing the YES softkey. If OVERRIDE LIMIT is not required press
the NO softkey and adjust dose above Soft Alert.
The bolus dose has been set above a Hard Limit. Check bolus setting and adjust to
appropriate required dose.
The patient weight has been set to a value which exceeds or is under a Soft Alert. Check
the weight setting, to continue press the
by pressing the YES softkey. If OVERRIDE LIMIT is not required press the NO softkey and
adjust the value within the limits.
The infusion rate has been set above a Hard Limit. Check infusion setting and adjust to
appropriate required rate.
1000DF00741 Issue 5
Alaris™ PK Plus Syringe Pump
b
button and then confirm
b
button and then confirm
b
button and then confirm
b
button and then confirm
b
button and then confirm
b
button and then confirm OVERRIDE LIMIT
28/52