Dynamic Monitoring System
Monitoring Options—General (Continued)
Select Monitoring Option (Continued)
4.
Press soft key for applicable option.
5.
Press ok soft key.
If Adjustable Pressure option is selected, pressure
system accuracy may be enhanced by ensuring that no
occlusion or other pressure source exists in
when activating
If Resistance option is selected, % Resistance is
displayed below bar graph while infusing.
If High Resistance option is selected, % Hi Resist. is
displayed below bar graph while infusing.
1.
If pressure limit adjustment is available, selection reads
Adjustable Pressure; otherwise, it reads Pressure.
2.
While the channel is on, the selected option, resistance alert, and
pressure alarm thresholds remain in effect until changed.
3.
Pressure alarm limits may be adjusted when operating in
Adjustable Pressure mode using the soft keys located below the
arrow symbols (see Adjustable Pressure Alarm on page 58).
4.
Maximum pressure limit settings may be configured by qualified
service personnel.
5.
Resistance alert limit may be adjusted using the soft keys located
below the arrow symbols (see Resistance Alert on page 54).
6.
High Resistance alert limit may be adjusted using the soft keys
located below the arrow symbols.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
1
2
3,4
.
/
RUN
HOLD
line
IV
5
6
Monitoring Options—General
OPTIONS
A B
Monitoring Options
A
Resistance
High Resistance
Adjustable Pressure
ok
OPTIONS
A B
Rate=
125.0 mL/h
A
VTBI= 1000.0 mL
VI=
0.0 mL
Primary
50% Resistance
OPTIONS
A B
Rate=
125.0 mL/h
A
VTBI= 1000.0 mL
VI=
0.0 mL
Primary
50% Hi Resist.
51