Dynamic Monitoring System
Monitoring Options—Resistance Mode

Detection of Downstream Occlusions

In the Resistance or High Resistance monitoring mode, a
RESISTANCE ALERT
resistance reaches the alert limit.
An
OCCLUSION DOWNSTREAM
of the following conditions occurs:
The measured resistance reaches
Resistance mode,
producing
mmHg per mL/h of flow. For the High
2
Resistance mode,
producing
mmHg per mL/h flow.
6
The configured pressure limit is exceeded. This limit may
be set, by qualified service personnel, from
mmHg (Pressure Limit, Maximum).
600
When a Downstream Occlusion is detected, one of the
following responses occurs:
If Auto Restart Plus feature is on, Checking Line message
is displayed and audible tone sounds.
If Auto Restart Plus feature is off,
alarm occurs.
DOWNSTREAM
52
condition occurs when the measured
condition is detected when one
100%
results from a resistance
100%
results from a resistance
100%
OCCLUSION
of scale. For the
mmHg to
1
Monitoring Options—Resistance Mode
A B
Rate=
125.0 mL/h
VTBI= 1000.0 mL
VI=
0.0 mL
Resistance Alert
A B
Rate=
125.0 mL/h
VTBI= 1000.0 mL
VI=
0.0 mL
Checking Line
A B
OCCLUSION
DOWNSTREAM
run
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
OPTIONS
A
OPTIONS
A
OPTIONS
A