Dynamic Monitoring System
Monitoring Options—Resistance Mode
Detection of Downstream Occlusions
In the Resistance or High Resistance monitoring mode, a
RESISTANCE ALERT
resistance reaches the alert limit.
An
OCCLUSION DOWNSTREAM
of the following conditions occurs:
•
The measured resistance reaches
Resistance mode,
producing
mmHg per mL/h of flow. For the High
2
Resistance mode,
producing
mmHg per mL/h flow.
6
•
The configured pressure limit is exceeded. This limit may
be set, by qualified service personnel, from
mmHg (Pressure Limit, Maximum).
600
When a Downstream Occlusion is detected, one of the
following responses occurs:
•
If Auto Restart Plus feature is on, Checking Line message
is displayed and audible tone sounds.
•
If Auto Restart Plus feature is off,
alarm occurs.
DOWNSTREAM
52
condition occurs when the measured
condition is detected when one
100%
results from a resistance
100%
results from a resistance
100%
OCCLUSION
of scale. For the
mmHg to
1
Monitoring Options—Resistance Mode
A B
Rate=
125.0 mL/h
VTBI= 1000.0 mL
VI=
0.0 mL
Resistance Alert
A B
Rate=
125.0 mL/h
VTBI= 1000.0 mL
VI=
0.0 mL
Checking Line
A B
OCCLUSION
DOWNSTREAM
run
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
OPTIONS
A
OPTIONS
A
OPTIONS
A