Electromagnetic Compatibility - CareFusion Alaris SE User Manual

Table of Contents

Advertisement

General Information
Epidural Administration (Continued)
WARNING
The Alaris SE pump can be used for epidural administration
of anesthetic and analgesic drugs. This application is
appropriate only when using anesthetics and analgesics
labeled for continuous epidural administration and
catheters intended specifically for epidural use. Use only
sets dedicated for use with the Alaris SE pump, without a
'Y' connector or injection port, for epidural infusions.
○ Epidural administration of anesthetic drugs: Use
indwelling catheters specifically indicated for short-term
(
hours or less) anesthetic epidural drug delivery.
96
○ Epidural administration of analgesic drugs: Use
indwelling catheters specifically indicated for either
short-term or long-term analgesic epidural drug
delivery.

Electromagnetic Compatibility

WARNING
Do not use the Alaris SE pump near Magnetic
Resonances Imaging (
Stereotaxis technology.
Do not use the Alaris SE pump near Therapeutic
Radiation equipment, such as Linear Accelerators.
Use of any
accessory, transducer, or cable
those specified might result in increased emissions or
decreased Alaris SE pump immunity.
CAUTION
Do not use the Alaris SE pump adjacent to or stacked
with other equipment. If adjacent or stacked use is
necessary, observe the Alaris SE pump to verify normal
operation in the configuration in which it will be used.
76
Warnings and Cautions (Continued)
) equipment, including
MRI
other than
User Manual—Alaris SE Pump, Rev2.X
Warnings and Cautions
Models 7100/7130, 7200/7230