Auto Restart Plus Feature

The Auto Restart Plus feature provides the ability to
automatically continue an infusion if downstream resistance or
pressure measurements indicate that an occlusion condition
has cleared within a
(excluding High Resistance Monitoring mode).
The Checking Line message and tone are presented when a
resistance measurement exceeds the alarm threshold of
If resistance measurements initiate the Checking Line
condition, the channel continues infusing in order to determine
if the measured flow resistance has changed. If the measured
flow resistance falls to any value below
seconds, the channel automatically resumes normal
40
operating conditions (excluding High Resistance Monitoring
mode).
Pressure measurements initiate the Checking Line period
when the pressure exceeds the configured limit. If the pressure
falls to less than one-third of the configured limit within
seconds, normal flow resumes. If the condition is not
40
cleared, the
OCCLUSION DOWNSTREAM
infusion stops until manually restarted.
This feature can be configured through the hospital
configuration settings to allow from
restarts. After the programmed number of restarts has
occurred or the
-second Checking Line period has been
40
exceeded, the channel immediately alarms
when resistance or pressure conditions
DOWNSTREAM
indicate an occlusion.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
-second Checking Line period
40
100%
to
1
9
100%
within
alarm occurs and
Checking Line
OCCLUSION
Monitoring Options—Resistance Mode
.
A B
Rate=
125.0 mL/h
VTBI= 1000.0 mL
VI=
0.0 mL
Primary
90% Resistance
OPTIONS
A
53