Programming
Drug Specific Dose Rate Calculator (DRC) (Continued)
\
Drug Name Not Listed (Continued)
6.
If there is a
value that needs to be cleared, press
VI
key or press
(zero) key, and then press
0
7.
To continue programming, press ok soft key.
8.
Verify that all parameters are correct and press run soft
key or
/
RUN
HOLD
9.
To briefly view current setup parameters from
page or during a running infusion, press
This is an optional feature that is configured by the facility.
Options are enabled through the instrument configuration
settings.
This feature allows an initial infusion rate to be set up for a
specific volume, automatically followed by a maintenance rate
(primary settings) from the same container. The primary
and
include the loading dose volumes. When the loading
VI
dose
reaches zero, a transition tone sounds (if transition
VTBI
tone feature is enabled), Load Dose Complete message is
briefly displayed, and the primary settings automatically take
effect.
Verify that the primary mode parameters are correct prior to
accessing the Loading Dose option.
1.
Select desired channel, as necessary. Channel must be
on hold in primary mode.
2.
Press
OPTIONS
32
Program (Continued)
key to start infusion.

Loading Dose

Program

key.
CLEAR
key.
ENTER
/
RUN
HOLD
soft key.
VTBI
WARNING
This mode is useful for delivering
fluid challenges. This feature is for
delivery from primary containers
only. Using this feature with two
separate containers might result in
unintended flow rates.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Loading Dose