General Information
Trumpet and Start-Up Curves (Continued)

Flow Characteristics Under Varying Delivery Conditions

Effects of pressure variations:
Under conditions of +
Alaris SE pump typically exhibits a long-term
accuracy offset of approximately -
values.
Under conditions of +
Alaris SE pump typically exhibits a long-term
accuracy offset of approximately -
values.
Under conditions of -
Alaris SE pump typically exhibits a long-term
accuracy offset of approximately -
values.
Resulting trumpet observation points typically
track those of accuracy; therefore, no significant
change in short–term variations result under these
pressure conditions.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
mmHg pressure, the
100
from mean
1.4%
mmHg pressure, the
300
from mean
1.5%
mmHg pressure, the
100
from mean
0.8%
Effects of negative solution container heights:
With a negative head height of -
Alaris SE pump typically exhibits a long–term
accuracy offset of approximately -
values.
Resulting trumpet observation points typically
track those of accuracy; therefore, no significant
change in short–term variations result under
negative head height conditions.
Effects of rate:
For applications where flow uniformity is a
concern, use of the Pressure Mode at rates of
mL/h or above is recommended.
1.0
Trumpet and Start-Up Curves
meters, the
0.5
from mean
5.8%
101