Programming; Primary Infusion - CareFusion Alaris SE User Manual

Table of Contents

Advertisement

References throughout this procedure to specific drugs and
drug doses are for illustration purposes only. Refer to specific
drug product labeling for information concerning appropriate
administration techniques and dosages.
IMPORTANT:
Prior to starting or restarting an infusion, verify that the
prescribed therapy is displayed.
Listen for audible tones.
Observe the flow rate in the IV tubing drip chamber to verify
the expected flow.
Displays: See Features and Displays on page 81 for a
description of the displays and soft keys, and instructions on
how to enter programming values.
Powering on and off: See Power On and Off on page 68 for
instructions on how to turn the instrument on and off.
Maintenance Reminder: See Respond to Maintenance
Reminder on page 70 for instructions on how to respond to a
Maintenance Reminder message.
1.
Perform "Start-Up" steps (see Start-Up on page 67).
Primary setup page is displayed.
Primary infusion rate is highlighted.
2.
Perform one of following steps:
If current primary infusion rate is appropriate, press
key.
ENTER
Enter new infusion rate and press
is highlighted.
VTBI
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230

Primary Infusion

key.
ENTER

Programming

WARNING
Prior to initiating a
verify that all displayed programming
parameters are correct.
A B
Rate=
0.1
mL/h
VTBI=
0.0
mL
VI=
0.0 mL
Primary
Limit=600 mmHG
,
/
RUN
HOLD
OPTIONS
A
19