References throughout this procedure to specific drugs and
drug doses are for illustration purposes only. Refer to specific
drug product labeling for information concerning appropriate
administration techniques and dosages.
IMPORTANT:
•
Prior to starting or restarting an infusion, verify that the
prescribed therapy is displayed.
•
Listen for audible tones.
•
Observe the flow rate in the IV tubing drip chamber to verify
the expected flow.
Displays: See Features and Displays on page 81 for a
description of the displays and soft keys, and instructions on
how to enter programming values.
Powering on and off: See Power On and Off on page 68 for
instructions on how to turn the instrument on and off.
Maintenance Reminder: See Respond to Maintenance
Reminder on page 70 for instructions on how to respond to a
Maintenance Reminder message.
1.
Perform "Start-Up" steps (see Start-Up on page 67).
•
Primary setup page is displayed.
•
Primary infusion rate is highlighted.
2.
Perform one of following steps:
•
If current primary infusion rate is appropriate, press
key.
ENTER
•
Enter new infusion rate and press
is highlighted.
VTBI
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Primary Infusion
key.
ENTER
Programming
WARNING
Prior to initiating a
verify that all displayed programming
parameters are correct.
A B
Rate=
0.1
mL/h
VTBI=
0.0
mL
VI=
0.0 mL
Primary
Limit=600 mmHG
,
/
RUN
HOLD
OPTIONS
A
19