CareFusion Alaris SE User Manual page 121

Table of Contents

Advertisement

Troubleshooting and Maintenance
Prompt
Program Lost Re-Enter Setting
Rate Out of Range
Set Must Be Loaded
Set Out
Set Pressure Baseline
Set Pri VTBI
Set Pri VTBI > Loading Dose
VTBI
Stop Timer to Change
Time Out of Range
Invalid Keypress During Programming:
The following Prompts might be seen if an invalid key is pressed during programming: Both A and B not
Running, Channel Not On, Complete or OK Setup, No Numeric Entries, Select Channel.
Invalid Keypress During Infusion:
During an infusion, if an invalid key is pressed, the following prompts might be seen: Dose Rate
Running, Loading Dose Running, Multi-Dose Running, Multi-Step Running, Pri Running, Sec Running,
or Timer Running (Multi-Dose program).
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Alarms, Alerts, Prompts (Continued)
Prompts (Continued)
A memory or power failure was
detected. Existing operating
parameters were erased.
Instrument has calculated a rate
less than
Flow Regulator segment is not
loaded in selected channel during
a manual pressure baseline
setting operation.
Pressure Baseline feature is on.
Flow Regulator segment is not
installed correctly.
Set Pressure Baseline was
selected in options mode.
A primary
programmed.
Loading dose
greater than primary
An invalid key was pressed while
timer was running in Multi-Dose
program.
Programmed step time exceeds
hours and
24
less than
Meaning
mL/h.
0.1
was not
VTBI
entered is
VTBI
.
VTBI
minutes, or is
59
minute.
1
Alarms, Alerts, Prompts
Response
Press continue soft key and
reenter all infusion settings.
Configurable options are not
affected.
Verify and reenter settings.
Load Flow Regulator segment in
selected channel. Repeat manual
pressure baseline setting.
Reinstall Flow Regulator segment.
Press ok soft key to set Pressure
Baseline or press return soft key to
go to exit.
Enter a primary
.
VTBI
Raise primary
or lower loading
VTBI
dose
, as appropriate.
VTBI
Wait several seconds for popup to
finish. Press stop timer soft key to
make changes.
Verify and reenter settings.
115

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

7100713072007230

Table of Contents