General Information
Downstream Occlusion
Time to Alarm
Bolus Volume
Environmental Conditions
Atmospheric Pressure
Relative humidity
Temperature Range
Fluid Ingress Protection
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Specifications (Continued)
Time to Occlusion and Bolus Volume data tested to standards defined in
AAMI ID26:1998, Section 51.101 b.
Time to Detect
Downstream Occlusion
(minutes)
Threshold Settings
Maximum
1 mL/h
Typical
Maximum
25 mL/h
Typical
When occlusion alarm pressure limit is set to maximum threshold setting, maximum
infusion pressure generated into a hard occlusion at 25 mL/h is 11.6±3.9 psi.
Bolus Volume Released
Upon Correcting
Downstream Occlusion
(mL)
Threshold Settings
Maximum
1 mL/h
Typical
Maximum
25 mL/h
Typical
Testing performed using Model 72003 administration set, at 68±8°F (20±4°C).
Operating
700 to 1060 hPa
20 to 90%
Noncondensing
41 to 104°F
(5 to 40°C)
IPX1, Drip Proof
Pressure
25 mmHg
600 mmHg
2
0.6
1
0.1
Pressure
25 mmHg
600 mmHg
0.5
<0.1
0.5
<0.1
Storage
500 to 1060 hPa
5 to 95%
Noncondensing
-40 to 140°F
(-40 to 60°C)
Specifications
Monitoring Options
Resistance and High
Resistance
100%
25 mmHg
75
2
30
0.6
25
1
1
0.1
Monitoring Options
Resistance and High
Resistance
100%
25 mmHg
0.5
0.5
0.3
<0.1
0.5
0.5
0.3
<0.1
100%
600 mmHg
7
4
3
1
100%
600 mmHg
0.5
<0.1
0.5
0.3
95