CareFusion Alaris SE User Manual page 101

Table of Contents

Advertisement

General Information
Downstream Occlusion
Time to Alarm
Bolus Volume
Environmental Conditions
Atmospheric Pressure
Relative humidity
Temperature Range
Fluid Ingress Protection
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Specifications (Continued)
Time to Occlusion and Bolus Volume data tested to standards defined in
AAMI ID26:1998, Section 51.101 b.
Time to Detect
Downstream Occlusion
(minutes)
Threshold Settings
Maximum
1 mL/h
Typical
Maximum
25 mL/h
Typical
When occlusion alarm pressure limit is set to maximum threshold setting, maximum
infusion pressure generated into a hard occlusion at 25 mL/h is 11.6±3.9 psi.
Bolus Volume Released
Upon Correcting
Downstream Occlusion
(mL)
Threshold Settings
Maximum
1 mL/h
Typical
Maximum
25 mL/h
Typical
Testing performed using Model 72003 administration set, at 68±8°F (20±4°C).
Operating
700 to 1060 hPa
20 to 90%
Noncondensing
41 to 104°F
(5 to 40°C)
IPX1, Drip Proof
Pressure
25 mmHg
600 mmHg
2
0.6
1
0.1
Pressure
25 mmHg
600 mmHg
0.5
<0.1
0.5
<0.1
Storage
500 to 1060 hPa
5 to 95%
Noncondensing
-40 to 140°F
(-40 to 60°C)
Specifications
Monitoring Options
Resistance and High
Resistance
100%
25 mmHg
75
2
30
0.6
25
1
1
0.1
Monitoring Options
Resistance and High
Resistance
100%
25 mmHg
0.5
0.5
0.3
<0.1
0.5
0.5
0.3
<0.1
100%
600 mmHg
7
4
3
1
100%
600 mmHg
0.5
<0.1
0.5
0.3
95

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

7100713072007230

Table of Contents