Troubleshooting and Maintenance
Qualified biomedical personnel may configure one of four
possible sensitivity levels. The instrument is also configurable
to permit the operator to clear (reset) any air registered in the
instrument's memory.
1.
To place channel on hold, press hold soft key.
2.
Remove air per hospital protocol.
3.
To resume infusion, reinstall set and then press
key.
HOLD
If air volume is clinically insignificant, press reset soft
key or
/
RUN
HOLD
key to resume infusion.
HOLD
Subsequent air bubbles trigger alarm.
Use this section in conjunction with the appropriate clinical
practice of hospital procedure.
When using the dual channel instrument, some messages also
display Channel A or Channel B, to indicate which channel is
affected. Always verify that the channel is selected before
making any changes.
ALARM—instrument or channel problem.
infusion stops
alarm bell icon illuminates
alarm tone sounds
rate LED display flashes
message appears in Main Display when channel is
selected
1.
Opening the latch or turning the channel off clears air memory.
108
Air-in-Line Assembly (Continued)

Air Bubble Detection

1
key, followed by run soft key or

Alarms, Alerts, Prompts

/
RUN
/
RUN
User Manual—Alaris SE Pump, Rev2.X
Air Bubble Detection
Models 7100/7130, 7200/7230