Regulations and Standards
Electromagnetic Environment (Continued)
Emissions Test
IEC 61000-4-2
Electrostatic
Discharge (ESD)
IEC 61000-4-4
Electrical Fast
Transient, Burst
(EFT)
(b)
IEC 61000-4-5
Power Line Surge
(b)
IEC 61000-4-8
Power Frequency
Magnetic Field (50/
60 Hz)
Compliance levels raised by
a.
b.
Performed at the minimum and maximum rated input voltage.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Compliance (Continued)
Electromagnetic Immunity
IEC 60601-1-2
Test Level
±
±
6 kV contact
±
±15 kV air
8 kV air
±
±
2 kV for power
supply lines
supply lines
±
±
1 kV for input/
output lines
output lines
±
±
1 kV differential
mode
mode
±
±
2 kV common
mode
mode
3 A/m
400 A/m 50 Hz
400 A/m 60 Hz
.
IEC 60601-2-24
Table 2
Compliance
Level
Floors should be wood, concrete, or
8 kV contact
(a)
ceramic tile.
(a)
If the floors are covered with synthetic
material, relative humidity should be at least
30%
If a connector testing exemption is used, the
following
adjacent to each connector.
Mains power quality should be that of a
2 kV for power
typical commercial or hospital environment.
1 kV for input/
Mains power quality should be that of a
1 kV differential
typical commercial or hospital environment.
2 kV common
Power frequency magnetic fields should be
(a)
at levels characteristic of a typical location
(a)
in a typical commercial or hospital
environment.
Electromagnetic Environment—
Guidance
.
sensitivity symbol appears
ESD
"Do Not Touch"
Compliance
121