General Information
Flow Rate Range
Ground Current Leakage
KVO Flow Range
Mode of Operation
Power Requirements
Rate Accuracy
Volume Infused Range
VTBI Range
96
Specifications (Continued)
0.1 to 600.0 mL/h in 0.1 mL/h increments (secondary mode)
0.1 to 999.9 mL/h in 0.1 mL/h increments (all other modes)
Electrical leakage current, enclosure: <100 microamperes
Electrical leakage current, patient: <10 microamperes
0.1 to 20.0 mL/h in 0.1 mL/h increments
Continuous
100-240 V~, 50/60 HZ (72 VA MAX), 3-wire grounded system
Class 1 with Internal Power Source
CAUTION
Variations of head height, back pressure, time, monitoring mode
option, pump tilt, or any combination of these might affect rate
accuracy. Factors that can influence head height and back
pressure are: Administration set configuration, IV solution
viscosity, and IV solution temperature. Back pressure might also
be affected by catheter type. See Trumpet and Start-Up Curves
on page 100
for data on how certain factors influence rate
accuracy.
For rates greater than 1 mL/h, up to 999.9 mL/h: ±5%, 95% of time with 95% confidence,
under conditions listed below.
For rates equal to or less than 1 mL/h: ±6.5%, 95% of time with 95% confidence, under
conditions listed below.
Rate Accuracy Test Conditions:
Infusion rate range: 0.1 to 999.9 mL/h
Head height: 24 ±1 in. (61±2.5 cm)
Test solution: distilled water
Environment temperature: 68±8°F (20±4°C)
Back pressure: 0 psi
Needle: 18 gauge
Set Model: 72003
Minimum collection volume: 6 mL
0.0 to 9999.9 mL in 0.1 mL increments
0.1 to 9999.9 mL in 0.1 mL increments (Basic Infusion, Drug Library Primary Infusion, IV
Fluid Infusion, and Multi-Step mode); 0.1 to 999.9 mL in 0.1 mL increments (all other
modes)
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Specifications