Flow Stop Test - CareFusion Alaris SE User Manual

Table of Contents

Advertisement

Getting Started
Check-In and Configuration (Continued)
Functional Test (Continued)
f.
Press
RUN/HOLD
instrument on hold.
g.
Release or open clamp and remove from tubing.
h.
Press
RUN/HOLD
should not reoccur.
9.
Perform Downstream Occlusion Test, as follows:
a.
Continue infusing (from step 8h).
b.
Verify that rate is set to
c.
Clamp off
IV
inches—to simulate a downstream occlusion.
d.
Allow instrument to run until it alarms
DOWNSTREAM
seconds.
60
e.
Press
RUN/HOLD
instrument on hold.
f.
Release or open clamp and remove from tubing.
g.
Press
RUN/HOLD
should not reoccur.
h.
Press
RUN/HOLD
1.
With an administration set primed and loaded in
instrument, turn power off.
2.
With all tubing clamps open and fluid container two or
more feet above instrument, verify that no fluid flows
through set.
3.
Open latch and remove set. Verify that no fluid flows
through set.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
key to silence alarm and put
key to resume infusion. Alarm
mL/h.
460
line just below instrument—about two
. Verify that this occurs within
key to silence alarm and put
key to resume infusion. Alarm
key to stop infusion.

Flow Stop Test

OCCLUSION
Check-In and Configuration
13