Troubleshooting and Maintenance
ALERT—possible change in infusion status.
•
channel continues to operate
•
alert tone sounds
•
message appears in Main Display
PROMPT—infusion status not changed.
Startup procedures were not completed or an invalid key
was pressed.
Alarm
Accumulated Air In Line
Air In Line
Battery Depleted (Plug In)
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Alarms, Alerts, Prompts (Continued)
Alarms
Air detector detected multiple
small bubbles.
Air detector detected an air
bubble larger than configured
threshold tolerance.
Battery is too low to operate
instrument.
Meaning
Alarms, Alerts, Prompts
Response
Press hold soft key. Open latch to
remove set. Clear air per hospital
protocol. Reinstall set. Press
to resume infusion.
HOLD
OR
If reset key is active and air bubbles
are clinically insignificant, press
reset soft key and then press run
soft key to resume infusion.
Ensure air-in-line sensors are
thoroughly cleaned.
Press hold soft key. Open latch to
remove set. Clear air per hospital
protocol. Reinstall set. Press
to resume infusion.
HOLD
OR
If reset key is active and air bubbles
are clinically insignificant, press
reset soft key and then press run
soft key to resume infusion.
At flow rates of
mL/h and below,
1.0
verify that upstream fluid path is
unobstructed.
Ensure air-in-line sensors are
thoroughly cleaned.
Plug power cord into an
immediately. Press run soft key or
to resume infusion.
/
RUN
HOLD
/
RUN
/
RUN
outlet
AC
109