Troubleshooting And Maintenance; Air-In-Line Assembly - CareFusion Alaris SE User Manual

Table of Contents

Advertisement

The Alaris SE pump Technical Service Manual is available
from CareFusion. It includes routine service schedules,
interconnect diagrams, component parts lists and descriptions,
test procedures, and other technical information, to assist
qualified service personnel in repair and maintenance of the
instrument's repairable components. Maintenance procedures
are intended to be performed only by qualified personnel.
Artifacts: It is normal for an infusion device to produce
nonhazardous currents when infusing electrolytes. These
currents vary proportional to the infusion device flow rate.
When an
monitoring system is not functioning under
ECG
optimal conditions, these currents might appear as artifacts,
simulating actual
ECG
abnormalities are caused by patient condition or the
equipment, place the Alaris SE pump on hold. If the
readings become normal, the
attention. Proper setup of the
these artifacts. Reference the appropriate
system documentation for instructions on setup and
maintenance.
The Air-in-Line Detection System provides clinicians the ability
to detect inappropriate amounts of air in the
instrument is configurable to allow single bubble and
accumulated air detection. Accumulated air detection is based
on measurement of the average percentage produced by small
air bubbles passing the detector.
Ensure that the tubing is properly inserted into the air detector
to avoid false alarms. The tubing may be reshaped to ensure
optimum contact with the sensors. Periodically clean the air-in-
line detector to ensure a clear signal can be received (see
Cleaning on page 116).
Air is detected by an emitter (air-in-line arm) which rotates into
position as the latch is closed. A receiver (air-in-line detector),
opposite the arm and just below the pumping mechanism,
sends the air-in-line information to the main processor.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
readings. To determine if
equipment requires
ECG
equipment should eliminate
ECG
ECG

Air-in-Line Assembly

Troubleshooting and Maintenance

ECG
ECG
ECG
monitoring
line. The
IV
Air-in-Line Arm
Air-in-Line Detector
107

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

7100713072007230

Table of Contents