General Information
Definitions (Continued)
Dynamic Monitoring
Auto Restart Plus
Monitoring Options
Trends
88
Configurable Settings (Continued)
Part of the Dynamic Monitoring system, and designed to help minimize nuisance
"occlusion downstream" alarms. It allows the instrument to automatically continue
an infusion following the detection of a downstream occlusion if downstream
pressure falls to an acceptable level within a 40-second Checking Line period. It
may be set to off (0 restarts) or to allow from 1 to 9 Checking Line restarts. If the
allowable number of restarts is exceeded, or when resistance or pressure conditions
indicate an occlusion, an occlusion downstream alarm occurs.
Dynamic monitoring allows one of the following monitoring modes to be selected:
Resistance mode, High Resistance mode, or Pressure mode. All of these modes
offer an optional Auto-Restart Plus feature and optional trend graph display.
Provides ability to display downstream pressure or flow resistance over time. Trend
graphs of 15 minutes, 1 hour, 4 hours, and 12 hours are available during normal
operation. When Trends is enabled and the instrument is operating in pressure
mode, a pressure trend graph is available. When Trends is enabled and the
instrument is operating in resistance mode, a resistance trend graph is available.
Configurable Settings
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230