Regulations and Standards
Electromagnetic Environment (Continued)
Emissions Test
IEC 61000-4-6
Conducted RF
IEC 61000-4-3
Radiated RF
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Compliance (Continued)
Electromagnetic Immunity—Life Support Equipment
IEC 60601-1-2
Test Level
3 Vrms
150 kHz - 80 MHZ
3V/m
80 MHZ - 2.5 GHz
Table 3
Compliance
(a)
Level
10 Vrms
Portable and mobile RF communications
equipment should be used no closer to the
Alaris SE pump (including cables) than
10 V/m
recommended separation distance
calculated from the equation applicable to
frequency of transmitter.
Recommended Separation Distance:
__ d = [-----] √P
__ d = [-----] √P 80 MHz - 800 MHz
__ d = [-----] √P 80 MHz - 2.5 GHz
d = recommended separation distance in
meters (m).
P = maximum output power rating of
transmitter in watts (W) according to
transmitter manufacturer.
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, should be less than compliance
level in each frequency range.
Interference might occur in the vicinity of
equipment marked with the following
symbol:
Electromagnetic Environment—
Guidance
12
V
2
12
E
1
12
E
1
(d)
Compliance
(b),(c)
(e),(f)
123