General Setup and Operation
Tamper Resist—Panel Lock Feature
The panel lock feature helps prevent unauthorized changes to
the instrument settings, including turning the instrument off.
The panel lock feature must be set to
configuration settings. To make changes or respond to an
alarm, the panel lock must be turned off. The panel lock key
( ) is located behind the handle.
The panel lock feature is always off when the instrument is
powered on because it must be off to power down the
instrument.
To turn panel lock feature on, press and hold
appears in lower display.
Panel Locked appears in Main
other key is pressed.
○ Dual channel instruments: When panel lock feature is
on, channel select and split screen keys can be used to
view settings.
To turn panel lock feature off, press and hold
disappears from lower display.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Start-Up (Continued)
in the hospital
ON
Display if any
LCD
until
GOLD
A
Rate= 125.0 mL/h
VTBI=1000.0 mL
VI=
until
GOLD
OPTIONS
B
A
0.0 mL
Panel Locked
Start-Up
69