Check-In and Configuration (Continued)
Rate Accuracy Qualification Test (Continued)
Test Setup
Burette Clamp
Equipment Stand
Table or Bench
1.
Fill solution container with clean tap water.
2.
Close AccuSlide flow regulator clamp on
then insert spike into solution container.
3.
Open AccuSlide flow regulator clamp and prime set.
Ensure that all air is expelled from set.
4.
Close AccuSlide flow regulator clamp.
5.
Connect output of set to one side of three-way stopcock.
6.
Load set into instrument and close latch.
7.
Verify no fluid flow or drops falling in drip chamber.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Burette (50 ml)
able or Bench
IV Tubing
Used Fluid Receptacle
Three-Way
Instrument Stand
Stopcock
set and
80VCS
Check-In and Configuration
Water Source
30 ±1 in
76.2
±2.5
cm
80VCS Calibration Set
5