Programming

Clear Volume Infused

The volume infused counter increments as fluids are infused
through a given channel. All fluids infused in primary mode, all
fluids infused in secondary mode, and all fluids infused in
mode are counted.
1.
To reset volume infused counter to
key.
2.
Press
or
CLEAR
This mode is designed to support automatic secondary
infusions (piggybacking) in the same channel. When the
instrument is programmed and delivering in the secondary
mode, the primary infusion is temporarily stopped and fluid is
drawn from the secondary container.
The primary infusion must be programmed and on hold before
a secondary infusion is programmed. The maximum rate for a
secondary infusion is
When the secondary
(if enabled), Secondary Complete message is briefly
displayed, and the primary infusion rate automatically resumes.
Delivery from the primary container also resumes when the
fluid level in the secondary line is level with the fluid in the
primary container.
For information regarding flow sensor use with secondary
infusions, see Flow Sensor on page 79.
22
Primary Infusion (Continued)
0.0
(zero) key and then press
0

Secondary Infusion

mL/h.
270
reaches zero, an audio tone sounds
VTBI
KVO
mL, press
soft
VI
key.
ENTER
Secondary Infusion
WARNING
Secondary applications require the
use of a check valve set on the
primary IV line.
The secondary solution container
must be higher than the primary
solution container.
The secondary VTBI settings
require consideration of variables;
such as, factory overfill and
medication additions.
Underestimating the volume causes
the remaining secondary solution to
be infused at the primary rate;
overestimating results in the primary
solution being infused at the
secondary rate. Multiple doses from
a single container are not possible.
The clamp on the secondary
administration set must be opened. If
the clamp is not opened, the fluid is
delivered from the primary container.
The secondary administration set
must be primed prior to beginning
the secondary infusion.
Verify that the secondary systems
are properly set up, to ensure that
there is a proper flow.
User Manual—Alaris SE Pump, Rev2.X