10-7-2 Leakage current limits

WARNING
CAUTION
Table 10-12:
Country
All (Except USA &
0.1 mA
Canada)
USA & Canada
0.1 mA
Table 10-13:
Country
All
LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Energy Control and Power Lockout for LOGIQ V3/V5/V5
Expert.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait for at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
6. Remove/disconnect the battery, if present.
Ultrasound System components may be energized.
Compare all safety-test results with safety-test results of
previously performed safety tests (e.g. last year etc). In case of
unexplainable abrupt changes of safety-test results consult
experienced authorized service personnel or GE for further
analysis.
The following limits are summarized for IEC 60601-1 Medical
Equipment Safety Standards. These limits are GEMS standards
and in some cases are lower than the above standards listed.
Chassis Leakage Current Limits - Accessible Metal Surface
Normal
Condition
Open Ground
0.5 mA
0.3 mA
Type BF Applied Part Leakage Current Limits - Probes Surface
Normal
Open
Condition
Ground
0.1 mA
0.5 mA
Electrical safety tests
Reverse
Polarity
0.5 mA
0.3 mA
Reverse
Open
Polarity
Neutral
0.5 mA
0.5 mA
Open Neutral
0.5 mA
0.3 mA
*Mains
Applied
5.0 mA
10-21