Flow TRIGGER Select SET and configure the parameter Tr. I = 0.3 l/min
Set PEEP to 5 cmH
Pressure
Select SET and configure the parameter Tr. I = -1 cmH
TRIGGER
Set PEEP to 5 cmH

5.3.11 Respiratory rates check

Set the following Respiratory Rate values and the corresponding Current Volume,
as shown in the table.
Make sure that the Respiratory Rate values displayed on the flow analyser and
on the ventilator monitored parameters fall within the corresponding acceptance
limits (+/- 10%).
Low/high
Note! Operations to be carried out during RR check
respiratory
Use the encoder knob and select - MENU - SETUP - ALARMS.
rate alarm
Set the values for Low RR alarms (e.g. 15bpm ) and High RR alarms (e.g.
50bpm ) so as to activate the visual and acoustic signals during respiratory
rate check phase.
generate a vacuum in the patient simulator
check if the Tr. I activates and initiates a breath
make sure that the device does not activate the trigger automatically
O
2
repeat the previously described tests and checks
generate a vacuum in the patient simulator
check if the Tr. I activates and initiates a breath
make sure that the device does not activate the trigger automatically
check the PAW value before and after activating the Tr. I = -1 cmH
O
2
repeat the previously described tests and checks
RR (bpm)
Current Volume
(ml)
12
500
40
100
70
50
O
2
Lower acceptance
limit (bpm)
10.8
36
63
O
2
Upper acceptance
limit (bpm)
13.2
44
77