Expiratory Flow Sensors Calibration - Siare Siaretron 4000 Service Manual

Icu ventilator, turbine-driven ventilation
Hide thumbs Also See for Siaretron 4000:
Table of Contents

Advertisement

6.2.2 Expiratory flow sensors calibration

Turn the knob and select the function:
1
Expiratory flow sensors calibration.
Siaretron displays the activation
screen for flow sensor calibration:
Press START to begin the
2
calibration.
Press ESC to verify the
calibration.
The activation of the Respiratory Flow Sensors Calibration program is necessary in the
following cases:
Noted differences are more than 15% ( over 100ml ) between the set Volume value ( VTi -
Vte ) and the expired Tidal Volume reading (Vte).
In case of first calibration of ventilator or replacement of flow sensors (INSP inside the
unit or EXP outside the unit), it is suggested to perform this Calibration.