Only use Integra approved power cords.
WARNING
Explosion Hazard - Do not use the CUSA Clarity System in the presence of flammable anesthetics or other volatile solvents.
WARNING
To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth.

10.6.3 Maximum Low Frequency Leakage

Touch current
Patient leakage current

10.7 Duty Cycle

Under maximum loading conditions, the CUSA Clarity console is suitable for ultrasonics activation times of 10 minutes on, 5 minutes off.
The system requires a minimum of one hour exposure at its operating temperature range before you use it.
See the EMC Information section, section 10.10 for further information on environmental conditions.
Electromagnetic Interference
The CUSA Clarity system console minimizes electromagnetic interference to other equipment used in the operating room. The system complies
with the requirements of IEC 60601-1-2: 2014 Fourth Edition.
CUSA® Clarity System Service and Maintenance Manual
AS3112, Straight
NBR 14136, Straight
GB2099, Straight
CEE 7/ XVII, Straight
CEI 23-50, Straight
JIS 8303, Straight
SANS 164-1, Straight
SEV Typ 12, Straight
BS 1363, 13A fused, Angled
<500µA
<50µA
Page 90 of 117