8.9 Status Bar Root Causes and Solutions

Status Bar root
causes & solutions
- Check if the footswitch connector is properly
connected.
Yes
Perform Setup Task:
- If the footswitch is already connected, replace the
Connect footswitch
footswitch (part number C7002).
- if the problem persists, call Integra for assistance.
No
- Check if the handpiece connector is properly
connected.
Yes
Perform Setup Task:
- If the handpiece is already connected, replace with a
Connect handpiece
new handpiece C7036.
- if the problem persists, call Integra for assistance.
No
- Check if the cartridge is properly installed and the door
Perform Setup
Yes
is closed.
Task: Connect
- if the problem persists, call Integra for assistance.
Tubing cartridge
No
Yes
- Check if the cartridge door is properly closed.
Close cartridge door
- if the problem persists, call Integra for assistance.
No
Yes
Perform Setup Task:
Prime handpiece
CUSA® Clarity System Service and Maintenance Manual
- Goto Setup Task Menu.
- Press Start Prime button.
No
Figure 8-4
Yes
Perform Setup Task:
Test handpiece
No
System is
Yes
overheating. Contact
service
No
Yes
System fan failure.
Contact service
No
Battery needs
Yes
replacement.
Contact service
No
End
Page 69 of 117
- Goto Setup Task Menu.
- Press Start Test button.
- Call Integra for assistance.
- Call Integra for assistance.
- The battery discharge is
considered complete but the
system is usable.
- Call Integra for assistance.
0645403-2-EN