Running the mass spectrometer at high flow rates

The ACQUITY UPLC system runs at high flow rates. To optimize desolvation,
and thus sensitivity, run the ACQUITY Xevo TQ-S system at appropriate gas
flows and desolvation temperatures. IntelliStart software automatically sets
these parameters when you enter a flow rate, according to the following table.
Flow rate versus temperature and gas flow:
Flow rate
(mL/min)
0.000 to 0.020
0.021 to 0.100
0.101 to 0.500
>0.500

Preparing the IntelliStart Fluidics system

For additional information, see
line" on page
Caution:
store large volume solvent reservoirs on top of the instrument.

Installing the reservoir bottles

Use standard reservoir bottles (30-mL) for instrument setup and calibration.
Use the Low-volume Adaptor Kit (included) to infuse smaller volumes. The
low-volume vials have a volume of 1.5 mL.
Required materials
Chemical-resistant, powder-free gloves
Source
temperature (°C)
150
150
150
150
Appendix D
B-22.
To avoid accidental spillage damaging the instrument, do not
Desolvation
temperature (°C)
200
300
500
0
and
"Connecting the liquid waste
Preparing the IntelliStart Fluidics system
Desolvation gas
flow (L/h)
800
800
1000
1000
2-5