System Messages; Status Messages And Displays - Sirona GALILEOS Service Manual

Hide thumbs Also See for GALILEOS:
Table of Contents

Advertisement

5 Messages

5.2 System messages

System code
Description
"System is starting"
S100
"Exposure not possible"
S110
"Sensor is prepaired (XX seconds)"
S150
Status displays
Easypad
(GALILEOS Comfort)
"Ready for exposure"
"X RAY"
"X RAY Active!"
"Please wait"
"Ready for exposure in XXs"
84
5.2
System messages
System messages
GALILEOS system messages
System codes are only displayed on the Multipad (GALILEOS Compact)
and the remote control. System messages are displayed in plain text on
the Easypad (GALILEOS Comfort).
GALILEOS system message table
5.3

Status messages and displays

Status messages and displays
On the control panel
Status displays: GALILEOS
Multipad
(GALILEOS Compact)
no special display; kV level and mAs
are displayed
LED lights up on Multipad.
Progress bar
XXs
On the Facescan
Facescan operating conditions
LED Blue (A)
Off
Blinking
On
On
On
Actions required
● Wait, no action required.
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure
● Wait, no action required.
The message will be deleted automatically (this
may take up to 10 minutes).
Description
System is ready for exposure.
Exposure in progress.
Unit waiting for operational readiness.
The cooling time countdown is running.
LED Green (B)
Meaning
On
Standby
Off
Ready for exposure
or
ready to send data
Off
Exposure
or
end of a firmware update via the
Facescan USB stick
Blinking
Data transfer
On
Boot process, firmware update or
reset to factory settings
Sirona Dental Systems GmbH
Service Manual GALILEOS
61 25 665 D3437
D3437.076.01.15.02 06.2012