Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS XG 3D / Ceph
A
T
R
63 03 452 D3352
D3352.031.05.08.02 10.2015
13.3
Checking and adjusting the touchscreen
1. Switch the unit on.
Once the unit software has started up, the start screen is
displayed on the touchscreen.
2. As soon as the start screen appears, press the 3 keys light localizer,
T, and R simultaneously (see (A)).
IMPORTANT: These keys must be pressed while the unit is booting
and the start screen is still displayed.
The first adjustment screen appears on the display.
You can abort the adjustment procedure at any time by pressing
"Abort"
the
3. Touch the top left corner of the screen.
The second adjustment screen appears on the display.
4. Touch the bottom right corner of the screen.
The third adjustment screen appears on the display.
5. Touch the center of the green square in the top right corner of the
screen.
IMPORTANT: Touch the green squares as close to the center as
possible so that the black dots that then appear will be positioned
exactly in the center of the squares.
The fourth adjustment screen appears on the display.
13 Adjusting and calibrating the unit
13.3 Checking and adjusting the touchscreen
button.
259