Therapy With Oxygen Supply - Weinmann VENTIlogic LS Instructions For Use Manual

Ventilogic ls - single patient circuit with patient valve; ventilogic plus - single patient circuit with patient valve
Hide thumbs Also See for VENTIlogic LS:
Table of Contents

Advertisement

Available languages

Available languages

Combination with an oxygen sensor and a humidifier
1. Connect the oxygen sensor directly to the therapy device.
2. Connect the bacteria filter to the outlet of the oxygen sensor.
3. Connect the humidifier to the outlet of the bacteria filter.
4. Connect the patient circuit to the humidifier.

4.7 Therapy with oxygen supply

4.7.1 Supplying oxygen
Warning!
• If oxygen is being supplied to the respiratory flow, smoking and naked flames
are forbidden. Risk of fire. The oxygen can accumulate in clothing, bed linen
or hair. It can only be removed by thorough ventilation.
• To supply oxygen, use only the therapy device connection provided for this
purpose. The therapy device has a safety valve which shuts off the oxygen
supply in the event of a fault, otherwise the oxygen supply cannot be stopped
automatically if a fault occurs. Never supply oxygen via masks or the T-adapter.
Note:
Supplying oxygen via a connection not intended for this purpose leads to an
erroneous volume display, as the oxygen flow cannot be included in the
measurement.
A supply rate of max. 15 l/min at < 1000 hPa pressure at the inlet for the oxygen supply is
permitted.
Use an oxygen sensor to monitor oxygen supply.
The oxygen source must have an independent flow regulation device. Ensure that you only
set the oxygen flow rate prescribed by the doctor on your oxygen supply device. It is
essential in this case to follow the safety instructions for handling oxygen as well as the
instructions for use of the oxygen device used.
Caution!
If the oxygen is humidified before being supplied, a tank with an overpressure valve
must be used, otherwise an overpressure will result in the event of faulty operation
which could lead to the humidification tank bursting or to the oxygen supply tubes
slipping off.
44
EN
Set up device

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ventilogic plus

Table of Contents