Connecting The Oxygen Supply - Weinmann MEDUVENT Standard Instructions For Use Manual

Hide thumbs Also See for MEDUVENT Standard:
Table of Contents

Advertisement

4.4 Connecting the oxygen supply

Hazardous therapy as a result of inadequate oxygen
concentration!
If oxygen is supplied from non-approved, inadequately cleaned or
damp oxygen sources, the pneumatic connections in the device
may be blocked by impurities or particles. This might injure the
patient.
 Use only oxygen sources which conform to the specifications
 Use only oxygen sources which are free of particles, clean and
Corrupted oxygen therapy resulting from use of unsuitable
oxygen!
Unsuitable oxygen might corrupt the therapy. This can cause
serious or life-threatening injury to the patient.
 Use only concentrator oxygen (90 % to 96 % oxygen) or
Failure of oxygen therapy resulting from inadequate supply of
oxygen!
An inadequate oxygen supply will prevent ventilation of the
patient. This might injure the patient.
 Check the pressure in the oxygen cylinder prior to ventilation.
Requirements
(see "16 Technical data", page
dry.
medical oxygen.
The oxygen cylinder has been filled.
The outlet flow of the inhalation source for the oxygen supply
is ≤ 15 l/min.
157).
MEDUVENT Standard
4 Preparation
EN
41

Advertisement

Table of Contents
loading

Table of Contents