Operation

Analog Output

Analog Output Edit

3-20
MIE ADR-1500 Instruction Manual
Note When the units /Mm are chosen, the RH correction becomes
automatically disabled.
From the Configure menu, press the
key to the Units edit screen.
UNITS

Units Edit

The Units Edit screen allows the user to change the option for units. Use
the ▲ and ▼ keys to toggle between the two unit options. The asterisk (*)
appears to the left of the field being edited.
From the Units screen, press the
Press the
UNITS
* ug/m3
The Analog Output screen indicates the status of the analog output
(enabled/disabled) and offers the user the option to enable or disable the
analog output and change the analog span range in corresponding units of
measure (ug/m
VDC for the analog output.
From the Configure menu, press the
key to the Analog Output screen.
ANALOG OUTPUT
The Units Edit screen allows the user to configure the analog output. The
values offered for selection are compatible with the selected units. The
values offered for /Mm (scattering coefficient) are: disabled, 10/Mm,
100/Mm, 1000/Mm, 10000/Mm, and 100000/Mm. The values offered for
3
μg/M
(concentration) are: disabled, 0.10 mg/m
3
mg/m
, 4.00 mg/m
1/Mm
key to save the units and return to the Units screen.
ENTER
1/Mm
3
or 1/Mm). The proportional voltage output range is 0–2
ENABLED
3
3
, 10.0 mg/m
, 40 mg/m
key and scroll the ▲ or ▼
ENTER
key to edit the units.
ENTER
key and scroll the ▲ or ▼
ENTER
3
, 0.40 mg/m
3
3
, 100 mg/m
, 400 mg/m
Thermo Fisher Scientific
3
, 1.00
3
, and