Using the Manual
Trigger
Using the SYNC
Output
Figure 3-11, Burst Run Mode Parameters
The manual trigger allows you to trigger or gate the 2074 directly
from the front panel. This button is active only when the generator is
placed in external trigger only. The MAN TRIG button is a second
function to the Enter button and can be used only when the display
is not in editing mode.
The 2074 has four sync outputs, one for every channel, located on
the rear panel. The outputs are always active, regardless if the main
output is turned on or off and therefore, for safety reasons, if you
are not sure if the output can damage your devices, disconnect the
sync outputs from your test gear immediately after you turn off the
generator. Each sync pulse can be programmed to have a different
position along the waveform length. The SYNC parameters are
shown in Figure 3-12. The Menu is accessible by selecting the
Outputs soft key as shown in Figure 3-3.
Figure 3-12, SYNC and Filter Parameters
Using the Instrument

Using the Manual Trigger

3
3-23