SYNC1, 2, 3 and 4

LAN

USB

GPIB

Figure ‎ 1 -6, 2074 Rear Panel
These BNC connectors generate synchronization signals, each for
every channel. The sync outputs ate active at all times, regardless
if the front panel main outputs were turned on or not and therefore,
if you leave cables hooked on these connectors, make sure they ca
not damage your equipment when you turn on the 2074.
This RG45 connector accepts standard Ethernet cable. Correct
setting of the IP address is required to avoid conflicts with other
instruments or equipment on the network. Information how to
change IP address and load instrument drivers to the computer is
provided in the Installation chapter of this manual.
This connector accepts standard USB-1 cable. The connection to
the host computer is automatic and does not require any address
setting from within the 2074. The first time the 2074 is connected to
the computer, it will request the driver file. This file is located on the
CD which is supplied with the instrument. Information how to install
the driver is provided in the Installation chapter of this manual.
This 24-pin connector accepts standard GPIB cable. The GPIB
address is configured using the front panel utility menu. The 2074
conforms to the IEEE-488.2 standard. Programming protocol is
SCPI version 1993.0. GPIB cables are available separately from
your Tabor dealer.