Sinewave Flatness

2074 Sine
Frequency
1 MHz
10 MHz
50 MHz
80 MHz
6-10
Table 6-8, Sinewave Spectral Purity
Spectrum Analyzer, Settings & Results
Start
Stop
1 MHz
100 MHz
10 MHz
200 MHz
10 MHz
200 MHz
Equipment: Oscilloscope
Preparation:
1. Configure the Oscilloscope follows:
Termination:
Setup:
2. Connect 2074 Channel outputs to the oscilloscope input
3. Configure the 2074 as follows:
Amplitude:
Output:
Frequency:
Test Procedure
1. Adjust the vertical controls of the Oscilloscope to get 6 division
of display
2. Perform Sine flatness tests on all channels using Table 6-9
Table 6-9, Sinewave Flatness
Error Limits
CH1
6 Divisions
6 ±0.3 Divisions
6 ±0.4 Divisions
6 ±0.4 Divisions
CH1
CH2
CH3
20 dB, 50 Ω feedthrough attenuator at the
oscilloscope input
As required for the test
6 V
On
Initially, 1 kHz then, as required by the test
Oscilloscope Reading
CH2
CH3
CH4
Reference
CH4
Pass
Fail
Pass
Fail