2074
User Manual
Test Results
USB Control
Test Results
LAN Control
6-24
2.
Connect 2074 CH1 output to the distortion analyzer input.
3.
Configure the 2074 as follows:
SCLK:
Waveform:
Output:
4.
Using ArbConnection prepare and download the following
waveform:
Wavelength:
Waveform:
Test Procedure
1.
Check the resulting trace on the oscilloscope
2.
Perform Sine wave distortion. It should be less than 0.1 %
Pass
Equipment: Distortion Analyzer, ArbConnection
Preparation:
1.
Set up the 2074 for USB operation and connect the instrument
to a host controller
2.
Connect 2074 CH1 output to the distortion analyzer input.
3.
Configure the 2074 as follows:
SCLK:
Waveform:
Output:
4.
Using ArbConnection prepare and download the following
waveform:
Wavelength:
Waveform:
Test Procedure
1.
Check the resulting trace on the oscilloscope
2.
Perform Sine wave distortion. It should be less than 0. 1 %
Pass
Equipment: Distortion Analyzer, ArbConnection
Preparation:
1.
Set up the 2074 for LAN operation and connect the
instrument to a host controller
2.
Connect 2074 CH1 output to the distortion analyzer input.
3.
Configure the 2074 as follows:
SCLK:
200 MS/s
Arbitrary
On
1 M points (2 M with an option installed)
Sine wave
Fail
200 MS/s
Arbitrary
On
1 M points (2 M with an option installed)
Sine wave
Fail
200 MS/s