2074
User Manual
Test Results
Period, Period
Averaged
Function
Generator Setting
Test Results
Pulse Width
6-26
2. Change the display time to Hold
3. Press the Reset/Arm button and verify that the frequency
reading is 120.00000 MHz, ±200 Hz
Pass
Equipment: Function Generator with at least 1 ppm accuracy
Preparation:
1. Configure the function generator as follows:
Frequency:
Wave:
Amplitude
2. Connect the function generator to the 2074 TRIG IN connector
3. Configure the 2074, as follows:
Auxiliary Function: Counter/Timer
Function:
Trigger Level:
Test Procedure:
1. Perform Period Accuracy tests using Table 6-14
Table 6-14, Period Measurement Accuracy
Error Limits
100.0 μs ±100 ns
10 kHz
10.00 μs ±100 ns
100 kHz
1.000 μs ±100 ns
1 MHz
2. Change the counter/timer function to Period Averaged
3. With the last function generator setting in Table 6-14, verify that
the period reading is 1.000000 μs ±10 ps
Pass
Equipment: Function Generator with at least 1 ppm accuracy
Preparation:
1. Configure the function generator as follows:
Frequency:
Wave:
Duty Cycle:
Amplitude
2. Connect the function generator to the 2074 TRIG IN connector
3. Configure the 2074, as follows:
Fail
As required by the test
Square
500 mV
Period
0 V
2074
Counter Reading
Pass
Fail
As required by the test
Square
As required by the test
500 mV
Fail