If the progress bar is not moving check the following for
possible problems:
1. If you are connecting to a LAN network, make sure your
device is connected with standard LAN wire to your wall
plug
2. If you use direct connection from your PC to the 2074,
your cable should be cross wired. You can get such cable
from any computer store near your area
3. If your network is using a managed switch, it is possible
that it is configure to break the package with broadcast
address and therefore, the only way to use NetConfig is to
connect the instrument directly to the PC with a cross
wired cable
Figure 7-5, Check for Progress Bar Movement
5.
As soon as the progress bar reached the far right hand of the
bar, click on the Refresh button. If your device was connected
and booted correctly, the device address will appear in the
device list, as shown in Figure 7-6.