User Manual
Recommended
Test Equipment
Equipment
Oscilloscope (with jitter package)
Digital Multimeter
Frequency Counter (Rubidium reference)
Function Generator (with manual trigger)
Accessories
Adjustment
Procedures
7-4
Recommended equipment for adjustments is listed in Table 7-1.
Instruments other than those listed may be used only if their
specifications equal or exceed the required minimal characteristics.
Also listed below are accessories required for calibration.
Table 7-1, Recommended calibration for Adjustments
Use the following procedures to calibrate the Model 2074. The
following paragraphs show how to set up the instrument for calibration
and what the acceptable calibration limits are.
Calibration is done with the covers closed and the 2074 connected
through an interface to a host computer. Any interface can be used
from the following: USB, LAN, or GPIB. Calibration requires that
ArbConnection utility be installed and interfaced to the instrument.
Calibration is performed from the Calibration Panel in ArbConnection.
To invoke this panel, one requires a password that is available to
service centers only. Contact your nearest Tabor service center for
information and permit to obtain your calibration password. Use the
following procedure to calibrate the generator:
1. Invoke ArbConnection
2. Click on the Calibration tab on the Panels bar
3. Expect to be prompted with the following dialog box
Model No.
LC684
2000
6020R
8020
BNC to BNC cables
50Ω Feedthrough termination
Dual banana to BNC adapter
Manufacturer
LeCroy
Keithley
Tabor Electronics
Tabor Electronics