Sync Position - Tabor Electronics 5064 User Manual

50 / 100 / 200 ms/s four channel arbitrary waveform generator
Table of Contents

Advertisement

2074
User Manual
Test Results
Test Results

SYNC Position

Test Results
Test Results
6-22
Pass
3.
Remove the cable from channel 1 and repeat the test on
channels 2, 3, and 4, using the appropriate sync output
connection on the rear panel for each channel
Pass
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
Amplitude:
2. Connect the 2074 CH1 output to the oscilloscope input (1)
3. Connect 2074 SYNC output CH1 to the oscilloscope input (2)
4. Configure model 2074 as follows:
Ch1 Output:
Test Procedure:
1. Verify trace on the oscilloscope shows synchronization pulses at
1 μs intervals. Verify that the synchronization pulse coincides
with the start of the waveform
2. Modify the SYNC Position parameters and observe that the
synchronization pulse position moves in reference to the
channel 1 output
Pass
3. Remove the cable from channel 1 and repeat the test on
channels 2, 3, and 4, using the appropriate sync output
connection on the rear panel for each channel
Pass
Fail
Fail
As required by the test
2 V/div
On
Fail
Fail