loops will the sequence step to the next assigned link.
To use the single advance mode, the 2074 must be in
programmed first to triggered run mode.
Mixed – This sequence advance mode allows combination of
automatic and stepped links in one sequence table. To use this
mode, in the sequence table, mark the Adv field "1" to flag stepped
link or "0" for continuous link. Then, download the sequence table to
the 2074. Note, to use this mode, the 2074 must be in continuous
operating mode. Step with a "0" bit assigned to a step it will
advance automatically to the next step. If "1" is assigned to a step,
the instrument will generate this step and its associated number of
repeats then, will wait for the next trigger to advance to the next
step.
1. Refer to Figure 3-20 and select the Advance Mode with the
appropriate soft key. The advance mode options, as shown in
Figure 3-20 will display. Use the dial or arrow keys to scroll
down to the required mode and press Enter to lock in the
selected mode.
Figure 3-20, Sequence Advance Options
The operating mode of the instrument, as selected from
the Run Modes menu, affects the way that the 2074 lets
you access the sequence advance mode parameter. If
you are in continuous mode, as shown in Figure 3-20,
you can select one of: AUTO, STEP, or MIXED. If you
already selected the triggered run mode, you'll be able to
select from AUTO, or SINGLE only as advance modes.