Test Results

Trigger Level

Test Results
Frequency:
Waveform:
Run Mode:
Output:
Test Procedure
1.
Toggle 2074 trigger slope from positive to negative visa versa
2.
Verify on the oscilloscope that the 2074 transitions are
synchronized with the slope of the trigger
Pass
Equipment: Oscilloscope, function generator
Preparation:
1.
Configure the Oscilloscope as follows:
Termination:
Setup:
2.
Connect 2074 Channel 1 output to the oscilloscope input
3.
Configure the function generator as follows:
Frequency
Run Mode:
Waveform:
Amplitude:
4.
Connect the function generator output to the 2074 TRIG IN
connector
5.
Configure the 2074 as follows:
Frequency:
Waveform:
Run Mode:

Trigger level:

Ch1 Output:
Test Procedure
1.
Verify that the 2074 outputs triggered waveforms spaced at 0.1
ms
2.
Modify the function generator offset to +2 V and change the
2074 trigger level to +4 V. Verify that the 2074 outputs triggered
waveforms spaced at 0.1 ms
3.
Modify the function generator offset to -2 V and change the
2074 trigger level to -4 V. Verify that the 2074 outputs triggered
waveforms spaced at 0.1 ms
Pass
Performance Checks
1 MHz
Sine wave
Triggered
On
Fail
20 dB, 50 Ω feedthrough attenuator at the
oscilloscope input
As required for the test
10 kHz
Continuous
Squarewave.
1 V
1 MHz
Sine wave
Triggered
0 V
On
Fail
6
Test Procedures
6-15